Traceable calibration of MPSS in fast-scan mode: Recommendations for PNSD measurement at urban traffic hot spots

Partículas en el camino
no eran contadas
pues eran muy finas
Aerosol metrology
UFP
Particle size distribution
MPSS

Arpit Malik, Johannes Rosahl, Sonja Pratzler, Jorge Saturno, Andreas Nowak, Traceable calibration of MPSS in fast-scan mode: Recommendations for PNSD measurement at urban traffic hot spots, Aerosol Science and Technology 1–20, doi: 10.1080/02786826.2026.2726892

Autores/as
Afiliaciones

Arpit Malik

Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Brunswick, Germany

Johannes Rosahl

Center for Atmospheric Research, University of Nova Gorica, Nova Gorica, Slovenia

Sonja Pratzler

Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Brunswick, Germany

Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Brunswick, Germany

Andreas Nowak

Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Brunswick, Germany

Fecha de publicación

octubre 2026

Doi
Otros detalles

Improvement of MPSS scan of ultrafine particles (UFP)

Abstract

To better represent real-world traffic emissions there is growing emphasis on measuring UFPs nearby traffic hotspots. Currently, the UFPs are monitored in Europe as PNSD measurements in the size range 10–800 nm using an MPSS. The associated quality assurance procedures were standardized by CEN/TS 17434:2020. The procedure prescribes validation of field deployed MPSS with a reference MPSS and a reference CPC over this full scan range using ambient aerosol. However, this wide scan range is suboptimal for traffic hotspots, where PNSDs peak at 50–70 nm with a negligible contribution of particles above 300 nm. Also, long scan time degrades temporal resolution, increasing the chance of missing high-emitter vehicles. Furthermore, the calibration using ambient aerosol (aged, stable charge states) may not adequately represent traffic hotspot’s fresh emission with unstable charge states. To address these gaps, our study puts forward an evolved calibration and validation procedure that incorporates MPSS operation in fast-scan mode (1-min and scan range 10–300 nm) and uses lab-generated soot particles as the test aerosols. To validate compliance with existing norms, a candidate MPSS (reference MPSS at PTB) is shown to be fulfilling the existing CEN criteria for a reference MPSS under this new procedure. Furthermore, an uncertainty evaluation approach is developed by compiling well-known and newly identified uncertainty sources. The evaluation suggests that MPSS measurement uncertainty can surpass CEN target limits under certain conditions (negative DMA voltage and low PN). Consequently, there is a need to comprehensively quantify these uncertainty sources, targeted correction algorithms and modified validation criteria.

Citation

@article{Malik2026,
  title = {Traceable calibration of MPSS in fast-scan mode: Recommendations for PNSD measurement at urban traffic hot spots},
  ISSN = {1521-7388},
  url = {http://dx.doi.org/10.1080/02786826.2026.2726892},
  DOI = {10.1080/02786826.2026.2726892},
  journal = {Aerosol Science and Technology},
  publisher = {Informa UK Limited},
  author = {Malik,  Arpit and Rosahl,  Johannes and Pratzler,  Sonja and Saturno,  Jorge and Nowak,  Andreas},
  year = {2026},
  month = Oct,
  pages = {1–20}
}